AI-6000KX Automatic Capacitance & Dissipation Factor test set
▪ Interference Suppression
▪ High Accuracy
▪ Safe Protection
▪ Multiple Functions
Tel:0531-88870011
Product Introduction
AI-6000KX Automatic 12kV Capacitance & Dissipation Factor Test Set is designed to measure dissipation or power factor in heavy electromagnetic interference environment such as in power plants or substations. Also it can be used in laboratory for high accurate test. The test set is all-in-one structure: including precision digital bridge, 12kV/200mA frequency conversion power unit, reference capacitor (Cn), and other electronic circuits. An external power supply can expand the test range. True portability is realized by the lightweight. Simple usage is realized by the full automation. High performance is realized by the digital process. AI-6000KX is equipped with megohm test function. The unit can test both C&DF and megohm for the same object without changing connections.
Product Features
▪ Interference Suppression:
The AI-6000K has advanced interference suppression circuit based on the frequency conversion and digital filter technique, the ratio of interference current to specimen current is 2:1(200%).
▪ High Accuracy:
Normally the accuracy of dissipation factor under no-interference environment is better than 0.00010.
▪ Safe Protection:
Output Protection, Input Protection, Grounded Protection, CVT, Fault Protection, No voltage overshoot, Pulse hint, etc
▪ Multiple Functions:
a)Basic test modes.
b)Use an external power supply.
c)Test at more higher voltage.
d)All test work will be done automatically after start the set.
Technical Parameter
C & tgδ | |
---|---|
Accuracy | Cx: ±( reading×1%+1pF ) |
DF tgδ: ±( reading×1%+0.00040 ) | |
PF cosφ: ±( reading×1%+0.00040 ) | |
Interference | Ratio of interference current to specimen current is 2:1 ( 200% ) |
Capacitance Range | Internal HV : 3pF~60000pF / 12kV, 60pF~1.2uF / 0.5kV |
External HV : 3pF~1.5uF / 12kV, 60pF~30μF / 0.5kV | |
Resolution : 0.001pF, 4 digitals | |
tgδ Range | No limit, Resolution 0.001% ( C/L/R specimen is automatic recognized ) |
Input current range | 10μA~5A |
Internal HV | 0.5~12kV / 200mA ( max ) |
Accuracy: ± ( reading×1%+10V ) | |
Control mode: Rise or fall smoothly | |
Frequency: 45Hz ~ 65Hz single frequency, 50±1Hz, 60±1Hz dual frequency | |
External HV | UST: 5A max, GST: 12kV/5A max, 38-72Hz |
Internal LV output for CVT mode | voltage range 3V~50V; current range 3A~30A |
CVT Ratio | Ratio Range: 10~99999, Ratio Accuracy: ±reading×1% |
Phase Range: 0~359.999°, Phase Accuracy: ±0.02° | |
Test Time | 30s typical ( may vary in different test mode ) |
Megohm | |
Voltage set | DC 100~10000V. Same as C&DF. HV terminal is positive. |
Accuracy | ±(Reading×2%+10V) when Rx>100Mhom. Test voltage will be reduced when test a smaller resistance(≤100Mhom). The min value is 45% of set value. |
Output Current | 100mA max. This is different from traditional meter for the current is very huge and danger. |
Range | 100kΩ ~ 1000GΩ |
Accuracy | 5%, 100kΩ~10GΩ when test voltage >250V |
5%, 10GΩ~100GΩ when test voltage >2500V | |
10%, 100GΩ~1000GΩ when test voltage = 10000V | |
Interference | ≤5mA, 50Hz / 60Hz |
Test mode | R (Megohm) 30~99 seconds adjustable |
AR (absorption ratio) = R60S/R15S | |
PI (polarization index) = R10MIN/R1MIN | |
Specimen | UST or GST |
Discharge | Quick discharge when test is finished |
Others | |
Power Supply | 95V~130V,180V~260V,50Hz/60Hz ( 1 phase power supply or generator ) |
Display | 128*64 dot matrix |
Printer | Thermal printer |
Communication Port | RS-232 |
Operating Temperature | -10℃~50℃ |
Storage Temperature | -20℃~60℃ |
Humidity | <90% non-condensing |
Dimensions | 34cm( L )*26cm( W )*27cm( H ) |
Weight | Main unit: 22kg , cables: 3.5kg |